the Principle of formation of an informative signal from a friction zone tribosoprjazhenija, consisting of spherical bodies, on an instance endoproteza

Interacting of components endoproteza represents process of chafing of rubbing surfaces of a head and a cup implantata which can lead in due course to formation of local defects, micro-and form macrodeviations, and also to origination vibroudarov, increasing thereby deterioration tribosoprjazhenija [17, 24, 42, 72].

At application of an electroresistance method not destroying control of a zone of a friction endoproteza is carried out by means of fluctuating electric resistance of contact of surfaces tribosoprjazhenija. As the control of a zone of a friction occurs in a dynamic regime electric resistance at tests will fluctuate and by that its analysis is expedient for spending, observing a measuring signal in the form of function from time Rt).izmeritelnyj signal Rt) as it was noted in gl. 2, contains the information on change of the contact square tribosoprjazhenija which formation is influenced by various external and internal factors (value and loading type, quality of made surfaces of components etc.). As natural process of traffic of a joint long-term (subsection 1.1.3) not destroying control is expedient for spending at the forced
Tests where the rotational speed of one of components endoproteza can be essential above real conditions. Thus, for reception of the measuring information from a friction zone it can be carried out by the simplified method observed more low.

In drawing 3.1 the block diagramme of reception of a signal from a zone of a friction of components endoproteza is presented.

1 - the current source; 2 - a head endoproteza; 3 - a cup endoproteza;

4 - the amplifier; 5 - a data collection pay

Drawing 3.1 - the Principle of measurement of frictional electric resistance

According to drawing 3.1 current source 1 affects a diagnostic current on tribosoprjazhenie, formed by a head 2 and a cup 3. Fluctuating resistance in a place of mating of components leads to an anode drop which amplitude is increased by the amplifier 4 and the pay of data collection 5 displays.

For possibility of the control of origination of local defects, vibroudarov, micro-and makrotkloneny the forms, the observed method should be added by the technical devices allowing:

- To test endoprotez in the conditions approached to the real;

- To control the force acting on endoprotez, at tests;

- To set traffic of one component concerning another, thereby to model chafing conditions implantata;

- To display parametres of an angle of turn for the control of change of the contact square tribosoprjazhenija, as consequence, the frictional
Resistance and anode drops at a complete revolution of one of components. In view of the aforesaid and the problems presented in section 3.1, the block diagramme of a developed method of diagnosing endoproteza will assume the following air (drawing 3.2) [68].

1 - the propeller; 2 - the turn angle-date trasmitter; 3 - the current source; 4 - a head;

5 - a cup; 6 - the electric channel of measurement; 7 - a data collection pay;

8 - the force sensing transducer; 9 - a lubricant layer; 10 - contact stains

Drawing 3.2 - the Structurally functional circuit design of installation of diagnosing

According to drawing 3.2 propeller 1 twirls a head endoproteza 4, and each change of a rule of a head 4 is displaid by the angle-date trasmitter of turn 2. Simultaneously from the current source 1 the diagnostic current on tribosoprjazhenie, formed by a head 4 and a cup 5 endoproteza moves. Owing to fluktuirovanija electric resistance at the expense of change of a lubricant layer 9 and the total square of contact stains 10 at interacting of surfaces the electric channel of the measurement consisting of amplifiers and the low-pass filter, 6 increases amplitude of an anode drop for its further machining. The control over acting force on endoprotez and character of loading occurs at the expense of the sensing transducer of force 8. In the end of test the measured parametres synchronised among themselves from the angle-date trasmitter of turn 2, the electric channel of measurement 6, the sensing transducer of force 8 are displaid by a gathering pay
Data 7 for postprocessing (transfer, scaling, interpretation of the gained values and representation of leading-outs).

<< | >>
Scientific source ZHIDKOV Alexey Vladimirovich. the ELECTRORESISTANCE METHOD And the DIAGNOSTIC DEVICE tribouzla with the spherical form of details (ON an instance endoproteza tazobedrennogo a joint with friction pair «metal - metal»). The dissertation on competition of a scientific degree of a Cand.Tech.Sci. An eagle, 2019. 2019

Other medical related information the Principle of formation of an informative signal from a friction zone tribosoprjazhenija, consisting of spherical bodies, on an instance endoproteza:

  1. THE LITERATURE LIST
  2. Chapter 2 Base, the program, a research technique
  3. THE LITERATURE LIST
  4. THE LITERATURE LIST
  5. the Central nervous system
  6. Design clinical kogortnogo pilot research of the teenagers subject to an emotionally-information strain.
  7. THE CONCLUSION
  8. THE CHAPTER I. THE LITERATURE REVIEW
  9. the mathematical software for the analysis the HORN of signals
  10. the literature List
  11. the literature List
  12. Structure of a method of secondary processing of the visualised information on condition PZH
  13. Model of formation of functional systems taking into account knacks and the got knowledge